| Electrical characteristics | |
| Current measurement ranges | Two ranges: ±10 A peak, ±60 A peak |
| Frequency range | DC to 40 kHz (−3 dB) |
| Power control | Power on when PicoBNC+ is connected |
| Resolution | 5 mA (with 20 kHz scope hardware filter and 500 Hz software filter) |
|
DC accuracy At 23 °C ± 5 °C, 70 %RH maximum When used with a PicoScope 4225A or 4425A oscilloscope |
0 to 60 A: ±4% of reading ± 630 mA 0 to 50 A: ±4% of reading ± 630 mA 0 to 20 A: ±2% of reading ± 190 mA 0 to 10 A: ±2% of reading ± 65 mA 0 to 5 A: ±2% of reading ± 35 mA 0 to 2 A: ±2% of reading ± 20 mA 0 to 1 A: ±2% of reading ± 20 mA 0 to 0.5 A: ±2% of reading ±20 mA |
| Zero offset adjustment |
Automatic on connection to PicoScope. Manual button in software. |
| Coupling to PicoScope | AC/DC, selectable in software. |
| Filtering | 20 kHz hardware lowpass filter in PicoScope. Enabled by default, selectable by user. |
| Remanence (0 A—60 A—0 A) | 300 mA typical |
| Working voltage | CAT II 300 V, pollution degree 2 |
| General characteristics | |
| Maximum conductor diameter | 9 mm (0.35") |
| Output connection | PicoBNC+® |
| Cable length | 2.85 m (9' 4") |
| Dimensions | 167 mm x 68 mm x 26 mm (6.6" x 2.7" x 1.0") |
| Weight | 0.25 kg (8.8 oz) |
| Environmental specifications | |
| Operating temperature range | 0 to 50 °C |
| Storage temperature range | −20 to +70 °C |
| Operating humidity | 70 %RH, max., noncondensing |
| Storage humidity | 80 %RH, max., noncondensing |
| Compliance | |
| Safety | EN 61010-1, EN 61010-2-032 |
| Environmental | RoHS, WEEE |